Appropriate for applications on skin. The foam has a surface treatment for a better affinity to adhesives and inks.
VITO MEDMOUNT 121 NT was designed for the manufacturing of multi-purpose electrodes for all applications involving ECG electrodes. The material is characterised by its exceptionally good adhesion and good skin compatibility.
The product must be stored unprocessed in the original carton at room temperature (20 ± 3 °C) and a relative humidity of 40 - 60 %. The shelf life is 12 months after receipt by the customer.
Peel adhesion on stainless steel
In acc. with DIN EN 1939
foam failure
Loop-Tack on polypropylene, according to FINAT FTM 9
≥ 8 N / 25 mm
Temperature resistance
-40 °C to 70 °C
Sterilisation feasibility
with steam, UV-light and gamma radiation